Medical wireless receiver
REC SW2.4LE-MED 5VDC PCB
Wireless receiver for medical applications
Circuit board for integration in medical device
1 digital validate output (relays)
1 serial interface (RS 232)
Technical data
General technical data
- Applied standards
- IEC 60601-1, ANSI/AAMI ES60601-1, IEC 62304, IEC 60529
- Antenna
- external antenna
- Mechanical life
- > 1 million operations
- Rated operating voltage Ue
- 5 VDC
- Rated operating current Ie
- not activated < 100 mA; activated < 200 mA
- Rated operating power Pe
- 500 mW
- Outputs
- 1 potential-free NO contact: K1 Reed-Relay
- Switching capacity
- max. 10 W
- Switching voltage
- 24 VAC/DC
- Switching current
- max. 0.5 A
- Conformity
- Global: cCSAus IEC 60601-1, -1-2, -2-22, -2-43
Europe: CE EN 300 328; EN 62479; EN 301 489-1; EN 62368-1
USA: FCC Title 47 CFR, Part 15
Canada: IC RSS-247, Issue 1
Japan: ARIB STD-T66
Wireless technology
- Frequency
- 2.4 - 2.4835 GHz
- Channel bandwidth
- 2 MHz
- Modulation type
- GFSK, adaptive frequency hopping of 40 channels
- Output power
- 3 dBm - 7 dBm
- Input sensitivity
- -93 dBm
- HF data rate
- 1 Mbps
- Interface data rate
- 115.2 kBd (UART)
- Wireless range
- 10 m (typical)
Note
Since this article is part of a medical product, it can only be validated in combination with the overall system. Proof of conformity to the Medical Device Regulation (MDR 2017/745) can therefore only be handled by the customer marketing the system. At the request of the customer, steute can affix a CE mark with the 4-digit number of the appropriate notified body.