Medical wireless receiver
REC SMF2.4-MED
Wireless receiver for medical applications
Wireless signal transmission per steute wireless SMF2.4-MED
1 serial interface (RS 232)
With back-up cable
Technical data
General technical data
- Applied standards
- IEC 60601-1, IEC 60601-1-2, ANSI/AAMI ES60601-1, IEC 62304, IEC 60529, ETSI EN 301 489-17
- Enclosure
- ABS, self-extinguishing UL 94 V-0, bone
- Antenna
- external antenna
- Degree of protection
- IP20 (IEC/EN 60529)
- Interfaces
- RS232
- Mechanical life
- > 1 million operations
- Operating voltage
- 5 V … 24 V DC
- Rated operating voltage Ue
- 24 VDC
- Rated operating current Ie
- < 500 mA
- Conformity
- Global: cCSAus IEC 60601-1, -1-2, -2-22, -2-43
Europe: CE EN 300 328; EN 62479
USA: FCC, Part 15C, Identifier: XK5-SMF241
Canada: IC RSS-247, Identifier: 5158A-SMF241
Wireless technology
- Frequency
- ISM 2.4 GHz band
- Channel bandwidth
- 2 MHz
- Modulation type
- GFSK, adaptive frequency hopping of 40 channels
- Output power
- 0 dBm - 7.5 dBm
- Input sensitivity
- -91 dBm
- HF data rate
- 2 Mbps
- Wireless range
- max. 30 m
Note
Since this article is part of a medical product, it can only be validated in combination with the overall system. Proof of conformity to the Medical Device Regulation (MDR 2017/745) can therefore only be handled by the customer marketing the system. At the request of the customer, steute can affix a CE mark with the 4-digit number of the appropriate notified body.